Research Article

Determination of the Chemical Stability of Cyanocobalamin in Medical Food by a Validated Immunoaffinity Column-Linked HPLC Method

Table 1

HPLC analysis condition for cyanocobalamin analysis.

ParametersConditions

ColumnAgilent Poroshell BONUS RP (4.6 mm × 150 mm, 2.7 μm)
Agilent Poroshell EC-C18 (4.6 mm × 50 mm, 2.7 μm)
Agilent Poroshell SB-AQ (4.6 mm × 150 mm, 2.7 μm)

Oven temperature40°C
Flow ratePreseparation column: 0.3 mL/min
Focusing/analytical column: 0.3 mL/min

Injection volume150 μL
Mobile phasePreseparation column: MeOH:5 mM KH2PO4 (15 : 85, v/v)
Focusing/analytical column: MeOH:5 mM KH2PO4 (25 : 75, v/v)

Wavelength361 nm