Abstract

A novel heart-cutting two-dimensional liquid chromatography coupled with tandem mass spectrometry (2D-LC-MS/MS) was developed for the qualitative and quantitative analysis of methomyl residue in tobacco. Compared to traditional methodologies, fairly high sensitivity and stability were achieved, and the sample procedure was simplified in the two-dimensional liquid chromatography (2D-LC) method. Although methomyl had poor retention performance in most of the reversed-phase liquid chromatography (RPLC) columns, an effective RP/RP strategy was successfully facilitated. An XB-Phenyl column was employed in the first dimension to effectively remove thousands of interference compounds in the matrix. In the second dimension, an ADME column was applied for further separation. After optimization of the separation conditions, a six-way valve was utilized for direct transformation of the target fraction from the 1st column to the 2nd column. A dynamic range of 2.5 ng/mL to 500 ng/mL was achieved with correlation coefficient (r2) greater than 0.9995. The limit of detection and limit of quantification were determined to be 0.69 and 2.30 ng/mL, respectively. The 2D-LC method shows high sensitivity, good reproducibility, and recovery for methomyl in tobacco samples. Therefore, the new method was quite suitable for routine analysis.

1. Introduction

Methomyl (S-methyl-1-N-[(methylcarbamoyl)oxy]thioacetimidate) is a systemic broad-spectrum carbamate pesticide, which is widely used in agriculture [1]. However, based on acetylcholinesterase inhibition, methomyl was toxic and hazardous to the environment and human beings [2, 3]. Consequently, it was classified as a high hazardous compound by the World Health Organization (WHO) [4]. Therefore, it is important to develop an efficient and sensitive analytical method for trace amount of methomyl residue in complex samples such as tobacco and tobacco products.

Many detection methods of methomyl were reported [414], such as liquid chromatography-tandem mass spectrometry (LC-MS/MS) [911], HPLC-fluorescence detection [12, 13], and so on. The LC-MS/MS method was widely used for higher sensitivity than the HPLC method. However, there was a serious complex matrix effect during detection of tobacco samples. One effective way was adding extra pretreatment steps to clean up samples. For instance, the QuEChERS (quick, easy, cheap, effective, rugged, and safe) method was used by adding two cleanup steps. However, for tobacco samples, matrix interferences were still serious after QuEChERS in the LC-MS/MS method, which affect ionization efficiency of methomyl. Furthermore, every 15–20 injections would show a visible decrease in intensity of MS. Therefore, a more convenient and efficient method was needed for measurement of methomyl.

Two-dimensional liquid chromatographic (2D-LC) technique provides more separation power and an extended dynamic range for complex samples [1517]. In our study, a RP/RP-2D-LC-MS/MS method was developed to determine methomyl residue in tobacco, which revealed fairly high selectivity and sensitivity. The new method was achieved by a combination of the XB-Phenyl column and the ADME column. In addition, a tandem spectrometry under a multiple reaction monitoring (MRM) mode was utilized for qualitative and quantitative analysis of methomyl. The approach was engaged for detection of a wide variety of samples.

2. Experimental

2.1. Apparatus and Chemicals

Two Agilent 1290 Infinity with diode array detector and 2 position/6 port switching valve were from Agilent Technologies Inc. (Agilent Technologies, Santa Clara, USA). AB SCIEX API 4000 Triple-Quad MS with electrospray ion source (AB SCIEX, USA) was utilized in this work.

Ammonium formate and methomyl standard were purchased from Sigma-Aldrich (Sigma-Aldrich, Saint Louis, USA). High performance liquid chromatography (HPLC) grade solvents including methanol (MeOH), acetonitrile (ACN), and formic acid were obtained from Merck (Merck, Darmstadt, Germany). Ultrapure water was obtained by using a Millipore Milli-Q water purification system (USA).

A QuEChERS kit including salt packets and SPE tubes was purchased from Agilent Technologies Inc. PTFE syringe filters (0.22 μm) were also obtained from Agilent Technologies. Tobacco samples were commercial products.

2.2. Preparation of Standard Solution

A 2.00 g blank tobacco sample (without methomyl) was weighted into a 50 mL centrifuge tube. The sample was shaken for 1 min after adding 10 mL water and 10 mL ACN. Then, the tobacco sample was pretreated by the simplified QuEChERS method: QuEChERS extraction salts were added, and then the tube was shaken for 2 min and centrifuged at 5000 rpm for 10 min. The obtained solution was filtered with a 0.45 μm PTFE syringe filter.

The 400 μL blank extraction was added to chromatographic bottles and dried up under nitrogen at room temperature. Then, calibration standards solutions were prepared in the chromatographic bottles. Concentrations of the standard solutions were 2.5, 5, 12.5, 62.5, 125, 250, and 500 ng/mL, respectively.

2.3. Sample Preparation

Methomyl was extracted from the tobacco sample and then cleaned up by a QuEChERS kit (Figure 1). 2.00 g of tobacco powder was weighted in a 50 mL centrifuge tube, 10 mL water was added, and the mixture was shaken for 1 min. Then, 10 mL ACN was added and the mixture was shaken for 2 min. After that, the extraction salt packet was added. The tube was shaken for 2 min and centrifuged at 5000 rpm for 10 min. The ACN layer was filtered and ready for 2D-LC-MS/MS analysis. For LC-MS/MS analysis, 1 mL of the ACN layer was added to the SPE tube for further cleanup. The tube was shaken for 2 min and centrifuged at 5000 rpm for 10 min. The upper layer was filtered and ready for LC-MS/MS analysis.

2.4. RP/RP-2D-LC-MS/MS Conditions

First-dimensional separation was as follows: Solvent A1, 0.1% formic acid with 10 mM ammonium formate in 5% MeOH; Solvent B1, 0.1% formic acid with 10 mM ammonium formate in 95% MeOH; 1D column, Welch Ultimate XB-Phenyl column (2.1 × 10 mm, 5 μm); column temperature, 30°C; injection volume, 3 μL.

Second-dimensional separation was as follows: Solvent A2, 0.1% formic acid with 10 mM ammonium formate in 5% MeOH; Solvent B1, 0.1% formic acid with 10 mM ammonium formate in 95% ACN; 2D column, Shiseido CAPCELL Core ADME column (2.1 × 150 mm, 2.7 μm); column temperature, 30°C.

Separation gradients of 1D and 2D chromatography are displayed in Table 1. Valve rotation procedure was as follows: 0–9.1 min, 1D column to waste; 9.1–11.9 min, 1D column to 2D column; 11.9 min, 1D column to waste.

The MS operation conditions were as follows: ion source, ESI; positive scan; MRM mode; ion spray voltage, 5500 V; ion source temperature, 500; curtain gas, 206.7 kPa; Gas 1, 414 kPa; Gas 2, 482.6 kPa; MS/MS acquisition time, 15–20 min. The MS analysis ions of methomyl are shown in Table 2.

3. Results and Discussion

3.1. The Heart-Cutting 2D-LC-MS/MS System Setup

During determination of methomyl in tobacco by the conventional LC-MS/MS, matrix interference was the major problem which seriously inhibits the signals of the target compound. To reduce the matrix effect, extra cleanup steps were required in the complex samples (Supporting information S1). However, tedious sample preparation steps not only extended the analysis time but also had bad effect on the repeatability and accuracy of the method.

In our work, a combination of two-dimensional liquid chromatography was applied to reduce the matrix effect and improve sensitivity of the analyte. The schematic diagram of the developed 2D-LC-MS/MS system is illustrated in Figure 2. The two-dimensional chromatography was coupled with a six-way valve. By the heart-cutting strategy, the interference compounds were effectively removed. Moreover, different stationary phases and mobile phases were employed to increase orthogonality of 2D-LC, and fairly high selectivity was achieved.

3.2. Optimization of 2D-LC Separation Conditions

Since retention of methomyl was poor on the C18 column, many other RPLC columns were tried and optimized. Consequently, a XB-Phenyl column and an ADME column were successfully applied in 1D and 2D, respectively. MeOH/water was used as a mobile phase in 1D, while ACN/water was employed in 2D for stronger eluting performance.

Heart-cutting was proceeded by the six-way valve shown in Figure 2. There were 3 stages in 2D-LC separation. (1) 1D to the waste: the valve was at position 1. 1D fraction was eluted from the 1D column to the waste. Meanwhile, the 2D column was balanced with the 2D mobile phase. (2) 1D to 2D: the valve was switched to position 2. The methomyl in 1D fraction was flowed to the 2D column. The flow rate of 1D was decreased to 0.2 mL/min, which was a benefit for retention of methomyl on the 2D column. In addition, methomyl was trapped on the 2D column. (3) 2D analysis: the valve was switched back to position 1. At 11.9 min, 2D separation started while other 1D fraction was eluted into waste at the same time. Finally, methomyl was determined qualitatively and quantitatively under MRM mode, and the MRM spectrum is shown in Figure 3.

Methomyl standard was utilized to find out retention time in the 1D separation (Figure 3(b)). The retention time of methomyl was between 9.6 and 11.4 min, so the heart-cutting time range was fixed between 9.1 and 11.9 min. As was shown in Figure 3(a), most of the interference compounds were removed by the heart-cutting strategy. There was almost no interference in determination of methomyl in tobacco samples (Figure 3(c)). Therefore, matrix effect was reduced significantly in MS after 2D separation, and relatively high sensitivity was achieved.

3.3. Method Validation

Method validation was studied in detection of methomyl by the 2D-LC-MS/MS method. The developed method showed a wide linear range (from 2.5 to 500 ng/mL) with correlation coefficient (R2) greater than 0.999. The limit of detection (LOD) and limit of quantification (LOQ) were calculated by signal to noise ratio (S/N). S/N of methomyl for LOD and LOQ was 3 and 10, respectively. LOD and LOQ of the method were 0.69 ng/mL and 2.30 ng/mL, respectively. The recovery of the method was investigated by spiking standard solutions with different concentrations (Table 3). The recovery of methomyl was between 93.1% and 108.2%. RSD of intraday repeatability ranged from 1.9% to 3.3%, and RSD of interday repeatability varied from 2.1% to 5.1%. Therefore, the new method exhibited good linearity, repeatability, and high sensitivity for methomyl analysis.

3.4. Comparison with Other Reported Methods

Table 4 summarizes the comparisons of the developed method and other reported methods for the determination of methomyl in different samples. The 2D-LC-MS/MS method revealed fairly high sensitivity and reproducibility. Moreover, the sample pretreatment steps were greatly simplified, which was also a benefit for the analyst and environment [921].

4. Conclusions

Detection of methomyl residue by conventional LC-MS/MS at extremely low level was difficult because of serious matrix interference in tobacco. A new strategy RPLC/RPLC-MS/MS was employed to effectively reduce the matrix effect. In the separation system, different columns and mobile phases were applied. After first-dimensional separation, most of the interference compounds were removed by heart-cutting. The methomyl fraction was transferred and separated on the second column.

The new method showed fairly high selectivity and sensitivity. LOD and LOQ for methomyl were 0.69 ng/mL and 2.30 ng/mL, respectively. Repeatability of intraday and interday was within 5.1%. In addition, sample procedure steps were simplified because sample purification and enrichment were achieved by the instruments. Furthermore, the system was running fully automatic. Therefore, the new method was quiet suitable for routine analysis. Also, the developed RPLC/RPLC-MS/MS system also provides potential application in analysis of other complex samples.

Data Availability

The data used to support the findings of this study are available from the corresponding author upon request.

Conflicts of Interest

The authors confirm that this article content has no conflicts of interest.

Authors’ Contributions

Shihao Shen and Min Chen contributed equally to this article.